Software Life Cycle for Medical Devices: IEC 62304, IEC 82304, and IEC 60601

These three standards are of particular importance during the implementation of the MDR requirements for software.

Contact
VDE Health

The European Medical Device Regulation (MDR) requires manufacturers to take the software life cycle of medical devices into account. If manufacturers wish to implement this requirement in practice, three standards are particularly important. In this article, we discuss what manufacturers must consider when addressing the software life cycle and how the standards specifically help in this regard.

Software life cycle for medical devices

A product’s life cycle describes the process from the initial product concept through to decommissioning or discontinuation of use. This concept also applies to software classified as a medical device. European legislation requires manufacturers to take these principles into account.

The MDR contains direct references to the life cycle of medical devices:

  • The manufacturer must keep the clinical evaluation and the associated documentation up to date throughout the product’s entire life cycle (Article 61, MDR).
  • The manufacturer must conduct continuous risk management throughout the product’s entire life cycle (Annex I, MDR).
  • The manufacturer must establish a quality management system and maintain it throughout the product’s entire life cycle (Annex IX, MDR).
  • As part of the type examination pursuant to Annex X of the MDR, the Notified Body reviews whether the product complies with the provisions of the Medical Device Regulation throughout its entire life cycle (Annex X, MDR).

Another provision explicitly refers to software as a medical device. It states:

  • “For products that include software as a component, or for products that are software in their own right, the software shall be developed and manufactured in accordance with the state of the art, taking into account the principles of the software life cycle, risk management - including information security - verification, and validation” (Annex I, MDR).

Implementation of the requirements for the software life cycle of medical devices

Annex I of the MDR defines the essential requirements for safety and performance of medical devices. If a manufacturer wishes to market software as a medical device, they must demonstrate that they meet these requirements.

The following standards are particularly helpful for practical implementation:

IEC 62304

When considering the software life-cycle process, the standard “Medical Device Software – Software Life-cycle Processes” (IEC 62304) is primarily relevant. The standard describes requirements for a life-cycle process and assigns specific activities and tasks to them.

It applies to the design and development of medical software and its maintenance. It does not matter whether the software itself is a medical device or whether it is used as an embedded or integral part of a medical device.

Application of the standard requires that the manufacturer develop and maintain its software as part of a quality management and risk management system.

The 2nd edition of IEC 62304 is still under revision. Following the first Committee Draft (CD1) in 2025, the responsible standards committee, IEC SC62A/MT49, is preparing a second Committee Draft (CD2). Until the second edition of the standard is published, the version IEC 62304:2006 + Amendment 1:2015 remains authoritative. The objective of the development of the standard is to create a uniform framework for all types of software. To this end, the scope is being expanded from “Medical Device Software” to “Health Software.” Overall, a distinction can be made between standalone software and device-associated software, each with a medical or (merely) health-related intended purpose, which operate on specific or general hardware platforms, respectively.

IEC 82304-1

Another standard, “Health Software - General Requirements for Product Safety” (IEC 82304-1), is also significant. It addresses issues related to safety and risks of software intended to be marketed independently of hardware (“stand-alone software,” “health apps”).

In this context, a health app can also be a medical device. The standard describes the requirements for manufacturers and covers the entire life cycle. This includes design and development, validation, installation, maintenance, and disposal.

Thus, its scope is broader than that of IEC 62304. The standard also describes overarching requirements that go beyond mere software development. There is also a clear focus on security.

IEC 60601-1

In addition to the two standards mentioned above, the standard “Medical Electrical Equipment - Part 1: General Requirements for Safety, Including Essential Performance Characteristics (IEC 60601-1)” plays a role as the foundational standard for medical electrical equipment.

Chapter 14 deals with “Programmable Electromedical Systems (PEMS)” and provides the general framework for a life-cycle approach to medical software.

What does IEC 62304 describe?

As a process standard, IEC 62304 describes five fundamental requirements for software processes:

  • Design and development
  • Maintenance
  • Risk management
  • Configuration
  • Problem solving

First, IEC 62304 requires manufacturers to classify the risks associated with their medical software. To this end, the standard specifies a three-class model consisting of safety classes A, B, and C. The safety classes are determined by the software’s contribution to a hazardous situation. Manufacturers apply the processes described above depending on the respective safety class.

The software development process

Once the software safety class has been determined, the software development is planned. Based on the software safety class, the manufacturer formulates a detailed software development plan consisting of tasks and activities, which must be kept up to date as development progresses.

The manufacturer then analyzes the software requirements. This includes, for example, characteristics such as timing behavior, operating system, memory size, default settings, interfaces, authentication, network protocols, and much more.

The manufacturer must then design suitable software architecture and subsequently describe how it intends to implement and integrate the software.

IEC 62304 explains the requirements for software architecture in detail. These include, for example, interfaces between components and specific requirements for “unknown” software components. The standard refers to such components as SOUP, “Software of Unknown Provenance,” or “off-the-shelf software.” This refers to generally available software that was not developed specifically for the medical device in question.

From a risk management perspective, SOUP naturally poses challenges. The development process also includes a description of software testing. The software development process concludes with information on documentation and archiving.

The software maintenance process

The software maintenance process begins with maintenance planning, which includes, for example, procedures for receipt, documentation, and traceability.

IEC 62304 then requires a systematic analysis of problems and changes that have arisen in connection with the software application. Among other things, this involves appropriate communication with users and the competent authorities. Finally, the standard describes how changes must be implemented and systematically approved.

The software risk management process

The software risk management process takes place within the framework of risk management according to ISO 14971 and first requires a hazard analysis. The focus is on components that could contribute to a hazardous situation, as well as their causes.

In the case of software, random errors do not occur, as they might in physical medical devices due to the poor quality of supplied and processed material or due to errors resulting from inattention on the part of production staff. If software contains an error, unlike a physical medical device, all copies of that version of the software are affected by this error. Software errors are therefore always systematic errors.

Here, too, SOUP plays a special role, as the standard explicitly requires an evaluation of publicly available lists of “SOUP anomalies.”

The hazard analysis is followed by a review of risk control measures, their Verification, and traceability documentation. Manufacturers should also explicitly consider and manage the risks associated with software changes.

The software configuration management process

IEC 62304 does not leave the “correct” configuration of medical software to chance. A detailed breakdown of identification, documentation, and approval steps ensures that the Manufacturer can find, adapt, and track the best possible configuration. The process, in turn, includes special consideration of SOUP.

The software problem-solving process

The standard requires that the manufacturer report, investigate, and notify relevant parties of problems arising in connection with the use of medical software. The manufacturer must also retain all records and perform a trend analysis of software problems. The process concludes with audit documentation.

What does IEC 82304-1 describe?

The standard first describes a series of requirements for healthcare software:

  • Risk management requirements (e.g., regarding the intended purpose and user profile),
  • User requirements (e.g., regarding information security and software support), and
  • system requirements (e.g., regarding interoperability and performance).

This is followed by a discussion of the software life cycle process. In this context, the standard refers directly to IEC 62304.

In addition, IEC 82304-1 describes the validation of health software and provides guidance to manufacturers on how to prepare the corresponding accompanying documents. The accompanying documents should address the following topics, among others:

  • User manual with a description of the health software
  • Safety warnings
  • Installation instructions
  • Start-up and shutdown procedures
  • Operating instructions
  • Notifications
  • Decommissioning
  • Technical description, e.g., for network integration

Finally, the standard sets out the framework for all activities performed by the Manufacturer after a health software product has been placed on the market. This includes appropriate maintenance and support of the software, as well as revalidation and orderly decommissioning at the end of the life cycle.

Defined processes are also important when it comes to documenting any errors or problems and communicating them to the competent authorities.

What does IEC 60601-1 describe?

In Chapter 14, IEC 60601-1 supplements the existing requirements for the risk management process and lifecycle assessment with necessary aspects of medical software (Programmable Electrical Medical Systems, PEMS).

This includes additions regarding the following PEMS aspects:

  • Documentation
  • Risk management Plan
  • Development lifecycle
  • Problem solving
  • Risk management process
  • Requirements specification
  • Architecture
  • Design and implementation
  • Verification
  • Validation
  • Modification
  • Integration into IT Networks

IEC 60601-1 proposes a development life cycle model for PEMS in Annex H. The model is based on the V-model for software development and is divided into a decomposition process and an integration process.

Starting with user requirements, the model describes, in a layered representation, various software Specification and development steps, which are subsequently combined for software verification and validation. The model relates the individual steps to Risk management.

In doing so, IEC 60601-1 extends the risk management process to include software-specific aspects, particularly regarding data, such as:

  • data availability,
  • data integrity,
  • data errors,
  • data timing, or
  • data security.

The integration of software into IT networks requires special precautions regarding safety and Risk management. IEC 60601-1 explicitly addresses this aspect. For example, the standard requires that the purpose of integration, the characteristics, and the configuration of the network be documented. In addition, the manufacturer must inform the software user of the risks arising from network integration.

Conclusion

If manufacturers of medical device software wish to comply with the statutory requirements of the MDR, the standard practice is to apply harmonized European standards.

The European Commission’s standardization mandate M/575 to CEN/CENELEC (last amended on May 27, 2024) continues to provide for the harmonization of IEC 62304 under the MDR and IVDR. To date, however, neither the current version of IEC 62304 (2006 + A1:2015) nor IEC 82304-1 has been published as a harmonized standard in the Official Journal of the EU. Both continue to be considered “only” as state of the art. This remains a disadvantage, particularly for manufacturers of standalone medical software.

We recommend that manufacturers of medical software familiarize themselves in detail with the practical application of the IEC 62304, IEC 82304-1, and IEC 60601-1 standards to comply with statutory requirements.


This article is intended solely for general informational purposes and does not constitute legal advice. All content was reviewed by experts prior to publication and is regularly checked for currency, completeness, and accuracy. Despite careful preparation and quality control, errors, omissions, or changes in the legal situation that have occurred in the meantime cannot be ruled out.